乳腺钼靶检测到的簇状微钙化的磁共振成像:有价值吗?

MR imaging of mammographically detected clustered microcalcifications: is there any value?

作者信息

Westerhof J P, Fischer U, Moritz J D, Oestmann J W

机构信息

Department of Diagnostic Radiology I, Georg-August-Universität, Göttingen, Germany.

出版信息

Radiology. 1998 Jun;207(3):675-81. doi: 10.1148/radiology.207.3.9609890.

Abstract

PURPOSE

To assess the value of dynamic magnetic resonance (MR) imaging in patients with mammographically detected suspicious microcalcifications.

MATERIALS AND METHODS

Sixty-three patients (age range, 31-84 years; mean age, 56 years) with mammographically suspicious clustered microcalcifications underwent dynamic MR imaging of the breast before surgical biopsy. MR imaging was performed with a 1.5-T unit and a two-dimensional fast low-angle shot sequence with a flip angle of 80 degrees. Thirty-two axial sections, 4 mm thick and without interval gaps, were acquired before and five times after administration of gadopentetate dimeglumine (dose, 0.1 mmol per kilogram of bodyweight). Histologic findings were used as the standard of reference. Any effect of MR imaging on surgical management was noted.

RESULTS

Biopsy findings verified five patients with T1 invasive carcinomas, 33 with ductal carcinomas in situ, 13 with proliferative fibrocystic disease, eight with nonproliferative fibrocystic disease, and four with sclerosing adenosis. Contrast material-enhanced dynamic MR imaging had a sensitivity of 45%, a specificity of 72%, and an overall accuracy of 56% in differentiating benign from malignant microcalcifications. All invasive tumors were detected with MR imaging. In no case was surgical management altered by MR imaging findings.

CONCLUSION

MR imaging of the breast is not reliable in differentiation of benign from malignant breast disease in mammographically suspicious clustered microcalcifications and has no effect on treatment.

摘要

目的

评估动态磁共振成像(MR)对乳腺钼靶检查发现的可疑微钙化患者的价值。

材料与方法

63例(年龄范围31 - 84岁,平均年龄56岁)乳腺钼靶检查发现可疑成簇微钙化的患者在手术活检前行乳腺动态MR成像。使用1.5T设备及翻转角为80度的二维快速低角度激发序列进行MR成像。在静脉注射钆喷酸葡胺(剂量为每千克体重0.1 mmol)前及注射后5次采集32层轴位图像,层厚4 mm,无层间距。以组织学结果作为参考标准。记录MR成像对手术治疗的任何影响。

结果

活检结果证实5例为T1期浸润性癌,33例为导管原位癌,13例为增生性纤维囊性疾病,8例为非增生性纤维囊性疾病,4例为硬化性腺病。对比剂增强动态MR成像在鉴别良性与恶性微钙化方面,敏感性为45%,特异性为72%,总体准确率为56%。所有浸润性肿瘤均被MR成像检测到。MR成像结果未改变任何一例的手术治疗方案。

结论

对于乳腺钼靶检查发现的可疑成簇微钙化,乳腺MR成像在鉴别乳腺良恶性疾病方面不可靠,且对治疗无影响。