Elliott C, Murray A
Regional Medical Physics Department, Freeman Hospital, Newcastle upon Tyne, UK.
Physiol Meas. 1998 May;19(2):159-64. doi: 10.1088/0967-3334/19/2/003.
The postural stability of 30 normal subjects in the age range 15 to 64 years was measured on five separate occasions, using the sway magnetometry technique, to determine the day-to-day repeatability of this measurement technique. On each occasion the pathlength and area enclosed by movement of the hips in a horizontal plane were measured over two 30 s periods with eyes open and eyes closed, with subjects standing on a firm base. Day-to-day repeatability had a standard deviation of 14% for pathlength and 42% for area. Neither the age nor the sex of the subject had any demonstrable effect on the results.