Dose rate constants for 125I, 103Pd, 192Ir and 169Yb brachytherapy sources: an EGS4 Monte Carlo study.

作者信息

Mainegra E, Capote R, López E

机构信息

Departamento de Física, Centro de Estudios Aplicados al Desarrollo Nuclear, Miramar, La Habana, Cuba.

出版信息

Phys Med Biol. 1998 Jun;43(6):1557-66. doi: 10.1088/0031-9155/43/6/014.

Abstract

An exhaustive revision of dosimetry data for 192Ir, 125I, 103Pd and 169Yb brachytherapy sources has been performed by means of the EGS4 simulation system. The DLC-136/PHOTX cross section library, water molecular form factors, bound Compton scattering and Doppler broadening of the Compton-scattered photon energy were considered in the calculations. The absorbed dose rate per unit contained activity in a medium at 1 cm in water and air-kerma strength per unit contained activity for each seed model were calculated, allowing the dose rate constant (DRC) A to be estimated. The influence of the calibration procedure on source strength for low-energy brachytherapy seeds is discussed. Conversion factors for 125I and 103Pd seeds to obtain the dose rate in liquid water from the dose rate measured in a solid water phantom with a detector calibrated for dose to water were calculated. A theoretical estimate of the DRC for a 103Pd model 200 seed equal to 0.669 +/- 0.002 cGy h(-1) U(-1) is obtained. Comparison of obtained DRCs with measured and calculated published results shows agreement within 1.5% for 192Ir, 169Yb and 125I sources.

摘要

利用EGS4模拟系统对192Ir、125I、103Pd和169Yb近距离放射治疗源的剂量测定数据进行了详尽的修订。计算中考虑了DLC - 136/PHOTX截面库、水分子形状因子、束缚康普顿散射以及康普顿散射光子能量的多普勒展宽。计算了每种种子模型在水中1 cm处介质中单位活度的吸收剂量率以及单位活度的空气比释动能强度,从而估算剂量率常数(DRC)A。讨论了校准程序对低能近距离放射治疗种子源强度的影响。计算了125I和103Pd种子源的转换因子,以便从用针对水的剂量校准的探测器在固体水模体中测得的剂量率获得液态水中的剂量率。得到了103Pd型号200种子源的DRC理论估计值为0.669±0.002 cGy h(-1) U(-1)。将所得DRC与已发表的测量和计算结果进行比较,结果表明192Ir、169Yb和125I源的一致性在1.5%以内。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验