Hederstedt B, Holm S E, Wadstrom T
J Clin Microbiol. 1978 Jul;8(1):50-3. doi: 10.1128/jcm.8.1.50-53.1978.
Comparison of the serum titers obtained with the Streptozyme, the antistreptolysin O, the antideoxyribonuclease B, and the antistreptohyaluronidase tests suggested that the Streptozyme test had failed to detect antibodies against streptococcal deoxyribonuclease B and hyaluronidase. Moreover, sera that were negative in the Streptozyme test could be shown by immunodiffusion to possess significant numbers of precipitins against extracellular factors produced by group A streptococci. Follow-up studies on patients with diagnosed streptococcal infections revealed elevated antideoxyribonuclease and streptohyaluronidase titers and increased numbers of precipitation lines without simultaneous increased titers by the Streptozyme test. There is thus a need for stricter control of possible batch-to-batch variations and more careful standardization of the antigen content of the Streptozyme test.
用链激酶试验、抗链球菌溶血素O试验、抗脱氧核糖核酸酶B试验和抗链球菌透明质酸酶试验所获得的血清滴度比较表明,链激酶试验未能检测出抗链球菌脱氧核糖核酸酶B和透明质酸酶的抗体。此外,链激酶试验呈阴性的血清经免疫扩散显示,含有大量针对A组链球菌产生的细胞外因子的沉淀素。对确诊为链球菌感染的患者进行的随访研究显示,抗脱氧核糖核酸酶和链球菌透明质酸酶滴度升高,沉淀线数量增加,而链激酶试验的滴度却没有同时升高。因此,需要更严格地控制可能存在的批次间差异,并更仔细地对链激酶试验的抗原含量进行标准化。