[在磁共振成像中使用造影剂注射器时的陷阱]

[Pitfalls when using a contrast media injector in MRI].

作者信息

Heverhagen J T, Hoppe M, Ishaque N, Froelich J J, Klose K J

机构信息

Abteilung für Strahlendiagnostik, Klinikum der Philipps-Universität, Marburg.

出版信息

Rofo. 1998 Aug;169(2):198-200. doi: 10.1055/s-2007-1015074.

Abstract

PURPOSE

Using a power injector to applicate Gd-DTPA we found a contrast enhancement of the pyelo-caliceal system even in the native studies, and hence we analysed pitfalls when using power injectors in MRI.

MATERIAL AND METHODS

We used a power injector Spectris (MedRad, Maastricht, Netherlands). In vitro artifacts were achieved by the mixture of contrast media and saline solution. We substituted contrast media by red water, NaCl by clear water.

RESULTS

Using power injectors in MRI, some pitfalls must be avoided, which can render investigations useless, especially dynamic contrast-enhanced investigations.

CONCLUSION

In our study we showed an easy way to overcome some pitfalls and use a power injector in MRI in a diagnostically helpful way. The simple use of valves inhibits the mixture of contrast media and saline solution.

摘要

目的

在使用高压注射器注入钆喷酸葡胺时,我们发现在平扫研究中肾盂肾盏系统也有强化,因此我们分析了在MRI中使用高压注射器时存在的问题。

材料与方法

我们使用了一台Spectris高压注射器(MedRad公司,荷兰马斯特里赫特)。通过混合造影剂和盐溶液来产生体外伪影。我们用红水替代造影剂,用清水替代氯化钠。

结果

在MRI中使用高压注射器时,必须避免一些问题,否则可能导致检查无效,尤其是动态对比增强检查。

结论

在我们的研究中,我们展示了一种简单的方法来克服一些问题,并以有助于诊断的方式在MRI中使用高压注射器。简单地使用阀门可防止造影剂和盐溶液混合。