Abstract

During the period from 1964 through 1994, the endolymphatic subarachnoid shunt operation was initially successful in eliminating endolymphatic hydrops and the symptoms and findings it produces in 76% of 645 ears of patients with Meniere's disease. After initial success, lasting from five weeks to nine years, endolymphatic hydrops suddenly returned due to obstruction of the Silastic shunt tube in 11% of patients. In these cases, prompt revision can often restore an initial good result. Histologic and immunologic examination of the material surrounding and occluding the tubes showed an allergic response to the Silastic material in most instances. Efforts to eliminate this cause of failure using a tube of new design and different plastic material are described.

摘要

在1964年至1994年期间,内淋巴蛛网膜下腔分流手术最初成功消除了梅尼埃病患者645只耳中76%的内淋巴积水及其产生的症状和体征。在最初成功的五周至九年后,11%的患者因硅橡胶分流管阻塞,内淋巴积水突然复发。在这些病例中,及时修复通常可恢复最初的良好效果。对阻塞分流管周围物质进行的组织学和免疫学检查显示,在大多数情况下,对硅橡胶材料存在过敏反应。本文描述了使用新设计和不同塑料材料的分流管来消除这一失败原因的努力。