[使用近红外光谱法评估体外循环期间的脑循环]

[Evaluation of cerebral circulation during cardiopulmonary bypass using near-infrared spectroscopy].

作者信息

Ohata T, Sawa Y, Ohtake S, Nishimura M, Hirata N, Kagisaki K, Taketani S, Yamaguchi T, Matsuda H

机构信息

First Department of Surgery, Osaka University Medical School, Japan.

出版信息

Jpn J Thorac Cardiovasc Surg. 1998 Jul;46(7):603-9. doi: 10.1007/BF03217788.

Abstract

Cerebral oxygenation level during cardiopulmonary bypass (CPB) was measured using near-infrared spectroscopy as a monitor of cerebral circulation in 30 patients. Six adult cases with thoracic aortic aneurysm were operated on using selective cerebral perfusion (SCP). CPB was established under moderate hypothermic temperature in 9 adult cases (hypothermic group, lowest blood temperature during CPB; 25 degrees C) and under normothermic temperature in 9 adult cases (normothermic group, lowest blood temperature during CPB; 34 degrees C). In congenital cases (n = 6), CPB was established under moderate hypothermic temperature (congenital group, lowest blood temperature; 25 degrees C). The oxyhemoglobin (HbO2) level showed a significant positive correlation with cerebral blood flow during SCP (r = 0.715). There was no significant correlation between SjO2 and HbO2 in the SCP group. The HbO2 levels in the hypothermic group after 30 and 60 min, from the initiation of CPB and 30 min. before the weaning of CPB were significant lower than the control level (p < 0.05). HbO2 levels in the congenital group after 0, 30 and 60 min. from the initiation of CPB and 30 min. before the weaning of CPB were significantly lower than the control level (p < 0.01). The deoxyhemoglobin (HbR) level in the hypothermic group after 30 and 60 from the initiation of CPB and 30 min. before the wearing of CPB were significantly higher than the pre level (p < 0.05). The mixed venous saturation (SvO2) in the normothermic group showed significant lower levels than those in the hypothermic group (p < 0.01). However, there was no significant difference in HbO2 levels between the two groups. In conclusion, these results suggest that near-infrared spectroscopy may be a noninvasive and useful technique for the cerebral circulation monitoring during CPB.

摘要

采用近红外光谱法测量了30例患者体外循环(CPB)期间的脑氧合水平,以此作为脑循环的监测指标。6例成年胸主动脉瘤患者采用选择性脑灌注(SCP)进行手术。9例成年患者在中度低温下建立CPB(低温组,CPB期间最低体温25℃),9例成年患者在常温下建立CPB(常温组,CPB期间最低体温34℃)。在先天性病例(n = 6)中,在中度低温下建立CPB(先天性组,最低体温25℃)。氧合血红蛋白(HbO2)水平在SCP期间与脑血流量呈显著正相关(r = 0.715)。SCP组中颈静脉血氧饱和度(SjO2)与HbO2之间无显著相关性。低温组在CPB开始后30分钟和60分钟以及CPB撤机前30分钟时的HbO2水平显著低于对照组水平(p < 0.05)。先天性组在CPB开始后0分钟、30分钟和60分钟以及CPB撤机前30分钟时的HbO2水平显著低于对照组水平(p < 0.01)。低温组在CPB开始后30分钟和60分钟以及CPB撤机前30分钟时的脱氧血红蛋白(HbR)水平显著高于术前水平(p < 0.05)。常温组的混合静脉血氧饱和度(SvO2)水平显著低于低温组(p < 0.01)。然而,两组之间HbO2水平无显著差异。总之,这些结果表明近红外光谱法可能是CPB期间监测脑循环的一种无创且有用的技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验