Butson M J, Yu P K, Metcalfe P E
City University of Hong Kong, Department of Physics and Materials Science, Kowloon Tong.
Phys Med Biol. 1998 Sep;43(9):2647-50. doi: 10.1088/0031-9155/43/9/015.
A radiotheraphy skin dose profile can be obtained with radiochromic film. The central axis skin dose relative to Dmax for a 10 x 10 cm2 field size was found to be 22%, 17% and 15.5% for 6 MV, 10 MV and 18 MV photon beams. Peripheral dose increased with increasing field size. At 10 MV the skin dose 2 cm outside the geometric field edge was measured as 6%, 10% and 17% for 10 x 10 cm2, 20 x 20 cm2 and 30 x 30 cm2 field sizes respectively. Off-axis skin dose decreased as distance increased from central axis for fields with Perspex block trays. For a 20 x 20 cm2 field, an approximately 5-8% drop in percentage skin dose was observed from central axis to the beam edge.