[Rapid Tendency Oriented Perimeter (TOP) with the Octopus visual field analyzer].

作者信息

Lachkar Y, Barrault O, Lefrançois A, Demailly P

机构信息

Institut du glaucome, Fondation Hôpital Saint-Joseph, Paris.

出版信息

J Fr Ophtalmol. 1998 Mar;21(3):180-4.

PMID:9759403
Abstract

PURPOSE

To study the capabilities of a new perimetric strategy with the Octopus visual field analyzer: the Tendency Oriented Perimetry.

METHODS

Tendency oriented perimetry (TOP) attempts to assess the visual field by using answers to questions to establish thresholds in the neighboring area. We evaluated 79 visual fields with the Octopus strategy using the program 32 and the TOP strategy split into the following groups: normal visual field or glaucoma suspects (52), moderately advanced glaucoma (16), advanced glaucoma (11). The following parameters were analyzed for the two strategies studied: examination time, MS (means sensitivity), MD (mean defect), LV (Loss Variance), short term fluctuation (RF) and the number of points with a deficit with different p values: p < 0.5; p < 1; p < 2; p < 5.

RESULTS

TOP perimetry showed a significant reduction of exploration time: 11.87 minutes with the Octopus 32 vs 2.49 minutes with the TOP strategy (p < 0.001). There is no significant modification for the other parameters (MD, MS, RF) except for the LV (Loss Variance) for the global analysis and for each separate group.

CONCLUSION

The TOP strategy reduces examination time significantly but seems to be less accurate especially for the calculation of the depth of each scotoma in comparison with the standard Octopus 32 perimetry.