Komersová A, Bartos M, Kalcher K, Vytras K
Department of Analytical Chemistry, University of Pardubice, Czech Republic.
J Pharm Biomed Anal. 1998 Apr;16(8):1373-9. doi: 10.1016/s0731-7085(97)00193-3.
Application of differential-pulse cathodic stripping voltammetry using a carbon paste electrode (consisting of carbon powder and liquid paraffin) have been investigated for trace determination of iron in 5-aminoisophthalic acid (AIPA). Samples were dissolved in 1 M HC1, pH was adjusted to 4-5 after addition of EDTA. Voltammetric measurements were performed after filtration. No sample decomposition (mineralization) was necessary. The method showed a good linearity between current and concentration from 3 x 10(-7) to 5 x 10(-5) mol dm-3 of iron, with a detection limit of 3 x 10(-7) mol dm-3 (resp. 1 ppm in solid AIPA). The results agreed well to those obtained by atomic absorption spectrometry (AAS) using electrothermic atomisation. For AAS measurement, however, microwave digestion of samples was necessary.