Evaluation of dual-tip micromanometers during 21-day implantation in goats.

作者信息

Reister C A, Koenig S C, Schaub J D, Ewert D L, Swope R D, Latham R D, Fanton J W

机构信息

Rothe Development Inc., San Antonio, TX, USA.

出版信息

Med Eng Phys. 1998 Sep;20(6):410-7. doi: 10.1016/s1350-4533(98)00049-6.

Abstract

Investigative research efforts using a cardiovascular model required the determination of central circulatory haemodynamic and arterial system parameters for the evaluation of cardiovascular performance. These calculations required continuous beat-to-beat measurement of pressure within the four chambers of the heart and great vessels. Sensitivity and offset drift, longevity, and sources of error for eight 3F dual-tipped micromanometers were determined during 21 days of implantation in goats. Subjects were instrumented with pairs of chronically implanted fluid-filled access catheters in the left and right ventricles, through which dual-tipped (test) micromanometers were chronically inserted and single-tip (standard) micromanometers were acutely inserted. Acutely inserted sensors were calibrated daily and measured pressures were compared in vivo to the chronically inserted sensors. Comparison of the pre- and post-gain calibration of the chronically inserted sensors showed a mean sensitivity drift of 1.0 +/- 0.4% (99% confidence, n = 9 sensors) and mean offset drift of 5.0 +/- 1.5 mmHg (99% confidence, n = 9 sensors). Potential sources of error for these drifts were identified, and included measurement system inaccuracies, temperature drift, hydrostatic column gradients, and dynamic pressure changes. Based upon these findings, we determined that these micromanometers may be chronically inserted in high-pressure chambers for up to 17 days with an acceptable error, but should be limited to acute (hours) insertions in low-pressure applications.

摘要

使用心血管模型的研究性研究工作需要确定中心循环血液动力学和动脉系统参数,以评估心血管功能。这些计算需要连续逐搏测量心脏四个腔室和大血管内的压力。在山羊植入21天期间,测定了八个3F双尖端微测压计的灵敏度和偏移漂移、寿命及误差来源。实验对象在左心室和右心室植入了成对的长期植入的充液接入导管,通过这些导管长期插入双尖端(测试)微测压计,并急性插入单尖端(标准)微测压计。急性插入的传感器每天校准,体内测量的压力与长期插入的传感器进行比较。长期插入传感器的增益校准前后比较显示,平均灵敏度漂移为1.0±0.4%(99%置信度,n = 9个传感器),平均偏移漂移为5.0±1.5 mmHg(99%置信度,n = 9个传感器)。确定了这些漂移的潜在误差来源,包括测量系统不准确、温度漂移、静水压力柱梯度和动态压力变化。基于这些发现,我们确定这些微测压计可在高压腔中长期插入长达17天,误差在可接受范围内,但在低压应用中应限于急性(数小时)插入。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验