Film digitizers are common devices in radiology departments involved with picture archive and communication systems (PACS) and teleradiology. In this paper, we studied the performance of film digitizers based on charge-coupled device detectors (CCD digitizers), and compared this with the performance of a laser digitizer (the de facto standard). Our focus was on the assessment of signal, noise and useful optical density range performance. A function (L* delta D) derived from the Rose model was used to evaluate these parameters in absolute terms, based their predicted ability to detect objects of specific size and optical density difference with respect to background. We studied CCD digitizers from four different vendors and found that none was able to reliably operate up to the maximum density of 3.0 required to digitize plain radiographs, while the laser digitizer was capable of this task. Our analysis also indicated that two of the four CCD digitizers were adequate for digitizing laser-printed cross-sectional images in certain cases. Finally, our analysis indicated that digitization of SMPTE pattern films along with visual assessment of the 5% and 95% contrast patches was not sufficient for determining the utility of film digitizers for clinical tasks. Computation of the L* delta D function provides a useful means of assessing the performance of film digitizers (e.g., for acceptance testing and quality control), and this technique may be adaptable for evaluation of other digital imaging modalities.
胶片数字化仪是放射科中用于图像存档与通信系统(PACS)和远程放射学的常见设备。在本文中,我们研究了基于电荷耦合器件探测器的胶片数字化仪(CCD数字化仪)的性能,并将其与激光数字化仪(事实上的标准)的性能进行了比较。我们重点评估了信号、噪声和有用光学密度范围性能。基于Rose模型推导的一个函数(LΔD)被用于从绝对意义上评估这些参数,这是基于它们预测检测特定尺寸和相对于背景的光学密度差异的物体的能力。我们研究了来自四个不同供应商的CCD数字化仪,发现没有一个能够可靠地运行到数字化普通X光片所需的最大密度3.0,而激光数字化仪能够完成这项任务。我们的分析还表明,在某些情况下,四个CCD数字化仪中的两个足以对激光打印的断层图像进行数字化。最后,我们的分析表明,对SMPTE图案胶片进行数字化以及对5%和95%对比度色块进行视觉评估,不足以确定胶片数字化仪用于临床任务的效用。LΔD函数的计算为评估胶片数字化仪的性能提供了一种有用的方法(例如用于验收测试和质量控制),并且这种技术可能适用于评估其他数字成像模态。