Malanoski S B, Belawski H, Horvath D, Smith W A, Golding L R
Mechanical Technology, Latham, New York 12110, USA.
ASAIO J. 1998 Sep-Oct;44(5):M737-40. doi: 10.1097/00002480-199809000-00089.
The Cleveland Clinic Foundation's Innovative Ventricular Assist System (IVAS) is distinguished by the use of a special hydrodynamic journal bearing to support the rotating assembly of the blood pump. In a permanently implanted blood pump, this bearing's characteristics of long life and high reliability are of paramount importance. In addition, this bearing's inherent self-pumping flow and the axial through flow caused by an imposed end-to-end pressure difference provides good washing, thus guarding against deposition. The basic computer analysis and preliminary testing results of this bearing were previously presented. This article reports the ongoing studies (both analytic and in vitro tests) on this innovative bearing as a component of the IVAS in general, with particular emphasis on its stable operating characteristics and reliability. The absence of vibration attributable to hydrodynamic instabilities related to the thick fluid film are both calculated and demonstrated during testing. A stable operating center of the rotor is shown to be inherent under magnetic side loads and resulting hydrodynamic bearing forces. A low shear as a result of large fluid-film thicknesses has been calculated, and low hemolysis has been shown by in vitro testing. Several unique design features of the bearing are believed to be responsible for this high level of performance.