Dose reduction in a paediatric X-ray department following optimization of radiographic technique.

作者信息

Mooney R, Thomas P S

机构信息

Northern Ireland Regional Medical Physics Agency, Forster Green Hospital, Belfast, UK.

出版信息

Br J Radiol. 1998 Aug;71(848):852-60. doi: 10.1259/bjr.71.848.9828798.

Abstract

A survey of radiation doses to children from diagnostic radiography has been carried out in a dedicated paediatric X-ray room. Entrance surface dose (ESD) and dose-area product (DAP) per radiograph were simultaneously measured with thermoluminescent dosemeters (TLDs) and a DAP meter to provide mean dose values for separate age ranges. Results of ESD and DAP were lower than the mean values from other UK studies for all ages and radiographs, except for the infant pelvis AP radiograph. Comparison of ESD and radiographic technique with CEC quality criteria highlighted a need for reduction of dose to infants and implied an increase in tube filtration might overcome the limitations of the room's three-phase, 12-pulse generator, allowing higher tube potentials to be used on infants. Additional tube filtration of 3 mmA1 was installed following assessment of dose reduction and image quality with test objects and phantoms, and confirmation from the paediatric radiologist that clinical image quality was not-significantly altered. The tube potential was increased from 50 to 56 kVp for the infant pelvis AP radiograph. The resulting ESD and effective dose fell by 51% and 38%, respectively. The CEC quality criteria have proved useful as a benchmark against which technique in X-ray departments can be compared, and as such are a useful tool for optimizing radiographic technique and reducing patient dose.

摘要

在一间专门的儿科X光室对儿童诊断性放射成像的辐射剂量进行了一项调查。使用热释光剂量计(TLD)和剂量面积积(DAP)仪同时测量每张X光片的入射表面剂量(ESD)和DAP,以提供不同年龄范围的平均剂量值。除婴儿骨盆前后位(AP)X光片外,所有年龄组和X光片的ESD和DAP结果均低于英国其他研究的平均值。将ESD和放射成像技术与欧洲电工标准化委员会(CEC)质量标准进行比较后发现,需要降低对婴儿的剂量,这意味着增加管滤过可能会克服该房间三相12脉冲发生器的局限性,从而能够在婴儿身上使用更高的管电压。在用测试物体和体模评估了剂量降低和图像质量,并得到儿科放射科医生确认临床图像质量未发生显著改变后,安装了3毫米铝当量的附加管滤过。婴儿骨盆APX光片的管电压从50千伏峰值提高到56千伏峰值。由此产生的ESD和有效剂量分别下降了51%和38%。事实证明,CEC质量标准可作为一个基准,用于比较X光科室的技术,因此是优化放射成像技术和降低患者剂量的有用工具。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验