Comparison of entrance surface doses of some X ray examinations with CEC reference doses.

作者信息

Ogunseyinde A O, Adeniran S A M, Obed R I, Akinlade B I, Ogundare F O

机构信息

Department of Radiology, University College Hospital, Ibadan, Nigeria.

出版信息

Radiat Prot Dosimetry. 2002;98(2):231-4. doi: 10.1093/oxfordjournals.rpd.a006715.

Abstract

Entrance surface dose (ESD) measurements have been carried out in Nigeria as part of the ongoing dose reduction programme. Thermoluminescence dosemeters (TLD) were used to measure skin entrance doses for four common radiographic views in three hospitals. The mean ESD for the PA chest examination in all the participating hospitals was in the range 0.12 - 4.46 mGy. The mean ESD for the AP skull. PA skull and LAT skull were 8.55, 5.17 and 6.97 mGy respectively. The mean ESD values are greater than the CEC reference doses, except for rooms 1 and 2 in UCH where the entrance surface doses for PA chest examination are below the CEC reference dose. The QA test results show non-compliance of the accuracy of tube voltage with acceptance limit in three rooms. The timer accuracy is also not within the acceptance limit in two rooms. The reproducibility of both the kVp and timer in all the rooms is good.

摘要

作为正在进行的剂量降低计划的一部分,尼日利亚开展了入射表面剂量(ESD)测量。使用热释光剂量计(TLD)测量了三家医院四种常见X光检查视图的皮肤入射剂量。所有参与医院中,PA胸部检查的平均ESD在0.12 - 4.46 mGy范围内。AP头颅、PA头颅和LAT头颅的平均ESD分别为8.55、5.17和6.97 mGy。平均ESD值高于欧洲共同体委员会(CEC)的参考剂量,但UCH的1号和2号房间除外,其PA胸部检查的入射表面剂量低于CEC参考剂量。质量保证测试结果显示,三个房间的管电压精度不符合验收标准。两个房间的定时器精度也不在验收范围内。所有房间的千伏峰值(kVp)和定时器的再现性良好。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验