视觉与器械比色评估。

Evaluation of visual and instrument shade matching.

作者信息

Okubo S R, Kanawati A, Richards M W, Childress S

机构信息

School of Dentistry, West Virginia University, Morgantown, W.V., USA.

出版信息

J Prosthet Dent. 1998 Dec;80(6):642-8. doi: 10.1016/s0022-3913(98)70049-6.

Abstract

STATEMENT OF PROBLEM

Visual color matching to determine shades in dentistry is inconsistent and unreliable. If accurate, instrumental measurement of tooth color would provide objective, quantified data to match natural teeth to clinical shade guides.

PURPOSE

This study evaluated and compared the ability of a new computerized colorimeter and a simple visual test to match ceramic shade guide teeth.

MATERIAL AND METHODS

Thirty-one (n = 31) observers with normal color vision were allowed unlimited time to match one set of Vita Lumin shade guide teeth to the corresponding shade guide teeth of a second Vita Lumin shade guide. The same test was administered to 14 of the observers several months later to determine within-subject variability. A computerized colorimeter (Colortron II) equipped with a positioning guide was used to measure the middle third of each shade guide tooth. Through a "match tool" present in the computer's software, readings from one shade guide were matched with readings of the other shade guide by using CIELAB measurements and DeltaE values. The mean number of correct matches by the colorimeter and of correct matches in visual test were compared with a 1-tailed t test. Repeatability for both tests was determined with a paired t test.

RESULTS

The Colortron II instrument correctly matched 8 of the 16 tabs (50% correct), whereas visual matching by examiners averaged 7. 7 of 16 correct matches (48% correct) (standard deviation 2.7). No statistically significant differences existed between the 2 methods. The colorimeter demonstrated 100% repeatability and the visual test demonstrated fair repeatability (correlation coefficient r =.60).

CONCLUSIONS

Shade determination by visual means was inconsistent. Accuracy of a new colorimeter in matching porcelain shade guide teeth was only slightly better.

摘要

问题陈述

在牙科领域,通过视觉进行颜色匹配来确定色度既不一致也不可靠。如果准确的话,牙齿颜色的仪器测量将提供客观、量化的数据,以便将天然牙齿与临床色度指南进行匹配。

目的

本研究评估并比较了一种新型电脑色度仪和一种简单视觉测试匹配陶瓷色度指南牙齿的能力。

材料与方法

31名色觉正常的观察者被给予不限时间,将一组维他全瓷比色板牙齿与另一组维他全瓷比色板的相应比色板牙齿进行匹配。几个月后,对其中14名观察者进行相同测试,以确定受试者内变异性。使用配备定位导向装置的电脑色度仪(Colortron II)测量每个比色板牙齿的中三分之一。通过计算机软件中的“匹配工具”,利用CIELAB测量值和色差(DeltaE)值,将一个比色板的读数与另一个比色板的读数进行匹配。通过单尾t检验比较色度仪正确匹配的平均数量和视觉测试中正确匹配的数量。用配对t检验确定两种测试的重复性。

结果

Colortron II仪器正确匹配了16个色片中的8个(正确率50%),而检查者的视觉匹配平均为16个中有7.7个正确匹配(正确率48%)(标准差2.7)。两种方法之间不存在统计学上的显著差异。色度仪显示出100%的重复性,而视觉测试显示出尚可的重复性(相关系数r = 0.60)。

结论

通过视觉手段确定色度是不一致的。一种新型色度仪在匹配烤瓷比色板牙齿方面的准确性仅略好一些。