Wachter S, Gerstner N, Colotto A, Battmann A, Gahleitner A, Haverkamp U, Pötter R
Universitätsklinik für Strahlentherapie und Strahlenbiologie Wien.
Strahlenther Onkol. 1998 Nov;174(11):589-96. doi: 10.1007/BF03038297.
Judgement of image quality and detail recognition of digitized and post-processed portal films presented on a computer monitor compared to the present standard, conventional portal films presented on a light box.
Conventional portal films of 3 different tumor sites (10 pelvis, 10 cranium, 10 vertebral column) were presented to a panel of 8 observers in 3 different matters: conventional film presented on a light box (Conv), digitized post-processed images (Dig-1) and digitized post-processed images (Dig-2) presented on a high resolution computer monitor. Subjective judgement of image quality, detailed recognition and time requirement of conventional films compared to monitor presentation were evaluated using a 5-scaled questionnaire (from 1 = much better to 5 = much worse). Furthermore the observers had to point out predefined anatomical bony structure on the conventional films (Conv) as well as on the digitized post-processed images (Dig-2). Standard deviations of the landmark outlined by 10 different observers were used as a criterion of objective detail recognition (Figure 1).
Image quality of digitized post-processed images presented on the computer monitor was judged statistical significant better than that of conventional films (pelvis 78%, vertebral column 62%, cranium 45% better) (Figure 3). Similar results were found for comparison of detail recognition: digitized post-processed images were scored better for pelvis in 81%, for vertebral column in 57%, for cranium in 40% (Figure 4, Table 1). Most benefit from portal film enhancement was found for pelvic images, where portal films are known to be of poor image quality (Figure 2). In contrast image quality of non-processed digital images compared to conventional films was graded worse (pelvis 69%, vertebral column 53%, cranium 71% worse) (Figure 4). Digital post-processed images were especially for the pelvis judged to require less time (pelvis 68%, vertebral column 26%, cranium 8% less time requirement) (figure 5). For the pelvis a statistical significant decrease of standard deviations was found for Dig-2 compared to conventional films, indicating an objective increase of image quality and detailed recognition (Table 2). In case of vertebral column and cranium no significant differences were evaluated (Table 3).
Digitized enhanced portal films presented on a computer monitor resulted in a quicker assessment and equal to better image quality as well as detail recognition compared to conventional films. Non-processed digitized images were judged to be of less image quality.
将计算机显示器上呈现的数字化及后处理门静脉造影图像的图像质量判断和细节识别与当前标准,即灯箱上呈现的传统门静脉造影图像进行比较。
向由8名观察者组成的小组以3种不同方式呈现3个不同肿瘤部位(10例骨盆、10例颅骨、10例脊柱)的传统门静脉造影图像:灯箱上呈现的传统胶片(Conv)、数字化后处理图像(Dig-1)以及在高分辨率计算机显示器上呈现的数字化后处理图像(Dig-2)。使用5级问卷(从1 = 好得多到5 = 差得多)评估与显示器呈现相比传统胶片的图像质量主观判断、细节识别和时间要求。此外,观察者必须指出传统胶片(Conv)以及数字化后处理图像(Dig-2)上预先定义的解剖骨结构。由10名不同观察者勾勒出的界标的标准差用作客观细节识别的标准(图1)。
计算机显示器上呈现的数字化后处理图像的图像质量经统计学判断显著优于传统胶片(骨盆部位好78%,脊柱部位好62%,颅骨部位好45%)(图3)。在细节识别比较中也发现了类似结果:数字化后处理图像在骨盆部位得分更高的比例为81%,脊柱部位为57%,颅骨部位为40%(图4,表1)。在门静脉造影图像增强方面,骨盆图像受益最大,已知该部位的门静脉造影图像质量较差(图2)。相比之下,未处理的数字图像与传统胶片相比图像质量被评为更差(骨盆部位差69%,脊柱部位差53%,颅骨部位差71%)(图4)。数字化后处理图像尤其被判定对于骨盆部位所需时间更少(骨盆部位所需时间少68%,脊柱部位少)。
与传统胶片相比,计算机显示器上呈现的数字化增强门静脉造影图像可实现更快的评估,且图像质量和细节识别相当甚至更好。未处理的数字化图像被判定图像质量较低。 26%,颅骨部位少8%)(图5)。对于骨盆部位,与传统胶片相比,Dig-2的标准差有统计学显著降低,表明图像质量和细节识别客观上有所提高(表2)。对于脊柱和颅骨部位,未评估出显著差异(表3)。