Measurement accuracy in confocal microscopy.

作者信息

Delorme R, Benchaib M, Bryon P A, Souchier C

机构信息

Université Claude Bernard, Laboratoire de Cytologie Analytique, Lyon, France.

出版信息

J Microsc. 1998 Nov;192(Pt 2):151-62. doi: 10.1046/j.1365-2818.1998.00423.x.

Abstract

Confocal laser scanning microscopy provides optical serial sections through thick biological samples, making it possible to perform both three-dimensional visualization and three-dimensional quantitative analysis. On human lymphocytes, we measured geometrical features, cell contents in DNA and in cyclin A and CDK1 proteins, localization and colocalization of these two proteins. Cells were acquired at a vertical sampling step of 0.5 micron, which gives sufficient information about cell labelling. For the purpose of obtaining fast and reliable data at a reduced time cost, we examined various possibilities to simplify acquisition. For example, it might be possible to increase the vertical sampling step to 2.0 microns while preserving an acceptable accuracy of measurements. Further limiting the acquisition to the central sections appeared to give only rough estimations about the whole cells. Finally, we compared confocal microscopy to conventional two-dimensional epifluorescence microscopy. Confocal microscopy appeared slightly less accurate as regards content estimation, but was an invaluable tool when investigating three-dimensional structures and, more especially, localization of proteins.