Schmit B D, Stellato T A, Miller M E, Mortimer J T
VA Medical Center, Cleveland, OH 44106, USA.
IEEE Trans Rehabil Eng. 1998 Dec;6(4):382-90. doi: 10.1109/86.736152.
Laparoscopic mapping of the phrenic nerve motor points using test stimulation was conducted for the implant of epimysial electrodes for diaphragm pacing in dogs. Both visual assessment of muscle activation and measurements of recruitment were useful for identifying an implant location resulting in a mean electrode placement approximately 14 mm from the phrenic nerve motor points in 16 dogs. Postmortem analysis of the stimulus test site locations and corresponding recruitment curves suggested that the phrenic nerve motor points could be predicted during the laparoscopic procedure to within 4.5 mm of the anatomical motor point.
在狗身上,使用测试刺激对膈神经运动点进行腹腔镜定位,以植入用于膈肌起搏的肌外膜电极。肌肉激活的视觉评估和募集测量对于确定植入位置都很有用,在16只狗中,电极平均放置位置距膈神经运动点约14毫米。对刺激测试部位位置和相应募集曲线的尸检分析表明,在腹腔镜手术过程中可以预测膈神经运动点,误差在解剖学运动点的4.5毫米范围内。