Larsimont V, Meins J, Fieger-Büschges H, Blume H
Zentrallaboratorium Deutscher Apotheker, Eschborn, Germany.
J Chromatogr B Biomed Sci Appl. 1998 Nov 20;719(1-2):222-6. doi: 10.1016/s0378-4347(98)00262-x.
A high-performance liquid chromatographic method for the determination of promazine in human plasma is described. The assay involves a single-step liquid-liquid extraction using pentane-2-propanol (98:2, v/v). The analyte of interest and the internal standard chlorpromazine were separated on a Spherisorb CN column using a mobile phase of acetonitrile-50 mM ammonium acetate (9:1, v/v). Electrochemical detection was achieved using an applied potential of +750 mV. The assay was validated according to international requirements prior to application to a pharmacokinetic study and was found to be specific, accurate and precise with a linear range of 0.25-25 ng ml(-1).