Suppr超能文献

用于COBAS AMPLICOR系统的AMPLILINK软件评估

Evaluation of AMPLILINK software for the COBAS AMPLICOR system.

作者信息

Kessler H H, Jungkind D, Stelzl E, Direnzo S, Vellimedu S K, Pierer K, Santner B, Marth E

机构信息

Molecular Diagnostics Laboratory, Institute of Hygiene, Karl Franzens University Graz, A-8010 Graz, Austria.

出版信息

J Clin Microbiol. 1999 Feb;37(2):436-7. doi: 10.1128/JCM.37.2.436-437.1999.

Abstract

The use of AMPLILINK version 1.0 software was evaluated for the operation and control of one COBAS AMPLICOR instrument and for two COBAS AMPLICOR instruments run simultaneously to perform and detect nucleic acid amplification reactions. A total of 3,384 results were analyzed. The initial accuracy of the results was 99.91%. Three errors of omission of transfer of data from the COBAS AMPLICOR to the AMPLILINK system were observed. Two of these errors were from a single specimen, where both the analyte and internal control results were not transmitted. These errors did not interfere with the correctness of any other data. There were no interruptions of runs, and no data were mixed. AMPLILINK increased convenience, saved labor, and was found to be a very useful addition for clinical laboratories performing molecular-diagnostic procedures with the COBAS AMPLICOR system.

摘要

对AMPLILINK 1.0版软件用于一台COBAS AMPLICOR仪器的操作与控制以及同时运行两台COBAS AMPLICOR仪器以进行和检测核酸扩增反应的情况进行了评估。共分析了3384个结果。结果的初始准确率为99.91%。观察到3例从COBAS AMPLICOR到AMPLILINK系统的数据传输遗漏错误。其中2例错误来自单个样本,该样本的分析物和内部控制结果均未传输。这些错误未干扰任何其他数据的正确性。运行没有中断,数据也没有混淆。AMPLILINK增加了便利性,节省了人力,对于使用COBAS AMPLICOR系统进行分子诊断程序的临床实验室来说,是一个非常有用的补充。

相似文献

1

本文引用的文献

5
Evaluation of the COBAS AMPLICOR MTB system.COBAS AMPLICOR 结核分枝杆菌检测系统的评估
J Clin Microbiol. 1997 Jun;35(6):1604-5. doi: 10.1128/jcm.35.6.1604-1605.1997.
9
International collaborative study on the second EUROHEP HCV-RNA reference panel.
J Virol Methods. 1996 Apr 26;58(1-2):175-85. doi: 10.1016/0166-0934(96)02011-3.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验