Evaluation of an ELISA test for determination of the serum transferrin receptor. Demonstration of discordance between results obtained with two methods.

作者信息

Virtanen M, Siimes M A, Krusius T, Pettersson T, Teppo A M, Viinikka L

机构信息

Department of Clinical Chemistry, University of Helsinki, Finland.

出版信息

Scand J Clin Lab Invest. 1998 Nov;58(7):561-7. doi: 10.1080/00365519850186184.

Abstract

We undertook this study to evaluate a recently introduced ELISA kit for determining serum transferrin receptor (TfR) concentration (TfR, Ramco Laboratories, Inc.), to produce reference values for healthy adults, and to compare the results with another commercially available reagent system. The mean (SD) recovery of added TfR was 88% (6%). In dilution studies, the ratio between the measured and expected values was 0.98 (0.11). The intra-assay and interassay coefficients of variation were from 5% to 7% and from 6% to 9% in a physiological and a supraphysiological concentration range, respectively, and from 13% to 16% in a subnormal concentration range. In healthy adults between 20 and 60 years of age, we observed no age- or sex-related differences in TfR values. Thus, the same reference interval, 3.0-8.2 mg l(-1), may be used for this population. The correlation between the results obtained with the Ramco TfR test and the Amgen Diagnostics Clinigen test was satisfactory (r=0.79). The Ramco TfR test produced higher values (Tf=0.40 (-0.45-1.25)+1.46 (1.16-1.75)* Clinigen). The number of samples that fell within the same concentration interval with both methods (low, normal or high in relation to the respective reference interval) was only 45% (27/60). The Ramco TfR test had fewer values falsely suggesting iron deficiency than the Clinigen test. Serum TfR methods need to be uniformly standardized.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验