Boyle N G, Anselme F, Monahan K M, Beswick P, Schuger C D, Zebede J, Josephson M E
Harvard Thorndike Electrophysiology Service, Cardiology Division, Beth Israel Hospital, Boston, Massachusetts 02215, USA.
Pacing Clin Electrophysiol. 1998 Dec;21(12):2685-7. doi: 10.1111/j.1540-8159.1998.tb00049.x.
Three cases of twiddler's syndrome in patients with an abdominally implanted cardioverter defibrillator (ICD) and an endocardial lead system are reported. The condition was detected when an increase in pacing threshold or lead impedance was noted at routine follow-up. Successful revision was possible in all patients; however, a recurrence in one patient necessitated removal of the device. Elderly female patients with some degree of obesity appear most susceptible. Minimizing pocket size and suturing of the device to the fascia should decrease the chance of developing this problem.
报告了3例植入腹部心脏复律除颤器(ICD)和心内膜导联系统的患者发生旋弄综合征的病例。该情况在常规随访中发现起搏阈值或导联阻抗增加时被检测到。所有患者均成功进行了修复;然而,1例患者复发,需要移除该装置。有一定程度肥胖的老年女性患者似乎最易患病。尽量减小囊袋大小并将装置缝合至筋膜应可降低发生此问题的几率。