Hydrodynamical comparison of aortic arch cannulae.

作者信息

Verdonck P R, Siller U, De Wachter D S, De Somer F, Van Nooten G

机构信息

Hydraulics Laboratory, Institute Biomedical Technology, University of Gent, Belgium.

出版信息

Int J Artif Organs. 1998 Nov;21(11):705-13.

PMID:9894746
Abstract

The high velocity of blood flow exiting aortic arch cannulae may erode atherosclerotic material from the aortic intima causing non-cardiac complications such as stroke, multiple organ failure and death. Five 24 Fr cannulae from the Sarns product line (straight open tip, angled open tip with and without round side holes, straight and angled closed tip with four rectangular, lateral side holes), and a flexible cannula used at the University Hospital of Gent (straight open tip) are compared in an in vitro steady flow setup, to study the spatial velocity distribution inside the jet. The setup consists of an ultrasound Doppler velocimeter, mounted opposite to the cannula tip in an outflow reservoir. An elevated supply tank supplies steady flow of 1.3 L/min of water. Exit forces at various distances from the tip are calculated by integrating the assessed velocity profiles. The pressure drop across the cannula tip is measured using fluid filled pressure transducers. The four sidehole design provides the lowest exit velocity (0.85 versus 1.08 m/s) and force per jet (0.03 vs 0.15-0.20 N). The round sideholes are useless as less than 1% of the flow is directed through them. Furthermore, the use of angled tip cannulae is suggested because the force exerted on the aortic wall decreases the more the angle of incidence of the jet deviates from 90 degrees. Pressure drop is the lowest for the 4 side hole design and highest for the open tip and increases when an angled tip is used.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验