Suppr超能文献

Extraction of pacemaker and implantable cardioverter defibrillator leads.

作者信息

Kantharia B K, Kutalek S P

机构信息

Division of Cardiac Electrophysiology, Hahnemann University Hospital, Philadelphia, PA 19102, USA.

出版信息

Curr Opin Cardiol. 1999 Jan;14(1):44-51. doi: 10.1097/00001573-199901000-00008.

Abstract

The common reasons for removal of pacing and defibrillator leads are infection, malfunction, or design defects such as fracture of J wires in Teletronics Accufix leads (Telectronics Pacing, Englewood, CO), which impose considerable risk for cardiac morbidity and mortality. Chronically implanted leads are fixed to the myocardium by fibrous tissue. Fibrous scar tissue may also encase the lead along its course. Furthermore, fragility of the lead and its tendency to break when extraction force is applied to overcome resistance imparted by the scar tissue add to the challenge of lead extraction. Thus, the extraction of chronically implanted leads is an important issue. Until a few years ago, the only methods available for the removal of chronically implanted leads were traction on the proximal segment of the lead and cardiac surgery. New techniques were developed to extract the leads by a transvenous approach using locking stylets, sheaths, snares, and retrieval baskets. Lead extraction using intravascular countertraction methods has since evolved as a specialty of its own. Progress has also been made in developing other system, such as Excimer laser energy for lead extraction. In this article, we discuss principles, techniques, and experience with these methods of extraction of chronic pacemaker and defibrillator leads.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验