Halperin R, Roa W, Field M, Hanson J, Murray B
Department of Oncology, Faculty of Medicine and Oral Health Sciences, University of Alberta, Edmonton, Canada.
Int J Radiat Oncol Biol Phys. 1999 Jan 1;43(1):211-6. doi: 10.1016/s0360-3016(98)00354-x.
Physiologic and non-physiologic tumor motion complicates the use of tight margins in three-dimensional (3D) conformal radiotherapy. Setup reproducibility is an important non-physiologic cause of tumor motion. The objective of this study is to evaluate and compare patient setup reproducibility using the reusable T-bar and the disposable expanded foam immobilization device (EFID) in radiation therapy for lung cancer.
Two hundred forty-four portal films were taken from 16 prospectively accrued patients treated for lung cancer. Patients were treated with either a pair of anterior and posterior parallel opposing fields (POF), or a combination of POF and a three-field isocentric technique. Each patient was treated in a supine position using either the T-bar setup or EFID. Six patients were treated in both devices over their treatment courses. Field placement analysis was used to evaluate 3D setup reproducibility, by comparing positions of bony landmarks relative to the radiation field edges in digitized simulator and portal images. Anterior-posterior, lateral, and longitudinal displacements, as well as field rotations along coronal and sagittal planes were measured. Statistical analyses of variance were applied to the deviations among portal films of all patients and the subgroup treated with both immobilization methods.
For the T-bar immobilization device, standard deviations of the setup reproducibility were 5.1, 3.7, and 5.1 mm in the anterior-posterior, lateral, and longitudinal dimensions, respectively. Rotations in the coronal plane and the sagittal plane were 0.9 degrees and 1.0 degrees, respectively. For the EFID, corresponding standard deviations of set up reproducibility were 3.6 mm, 5.3 mm, 5.4 mm, 0.7 degrees and 1.4 degrees, respectively. There was no statistically significant difference (p = 0.22) in the 3D setup reproducibility between T-bar and EFID. Subgroup analysis for the patients who were treated with both immobilization devices did not reveal a difference either. There was no consistent systematic error from simulator to treatment unit identified for either immobilization device.
Although the optimal immobilization technique and patient positioning for thoracic radiotherapy have yet to be determined, this study indicates that T-bar is comparable with EFID in its setup reproducibility. In view of the inherent advantages of T-bar, it has become a standard immobilization device at our institution. The observed range of displacements in field positioning with either immobilization device implies that one cm (two standard deviations [SD] of setup error) will be a more appropriate margin to allow for setup variability in radiation therapy for lung cancer.
生理和非生理性肿瘤运动使三维(3D)适形放疗中使用紧密边界变得复杂。摆位重复性是肿瘤运动的一个重要非生理性原因。本研究的目的是评估和比较在肺癌放射治疗中使用可重复使用的T型杆和一次性发泡固定装置(EFID)时患者的摆位重复性。
对16例前瞻性入组的肺癌患者拍摄了244张射野片。患者接受一对前后平行相对野(POF)治疗,或POF与三野等中心技术联合治疗。每位患者仰卧位使用T型杆摆位或EFID摆位。6例患者在整个治疗过程中使用了两种装置。通过比较数字化模拟机和射野片图像中骨标志相对于辐射野边缘的位置,采用射野放置分析来评估3D摆位重复性。测量前后、左右和纵向位移,以及沿冠状面和矢状面的射野旋转。对所有患者的射野片之间的偏差以及使用两种固定方法治疗的亚组进行方差统计分析。
对于T型杆固定装置,摆位重复性的标准差在前后、左右和纵向维度分别为5.1、3.7和5.1mm。冠状面和矢状面的旋转分别为0.9度和1.0度。对于EFID,相应的摆位重复性标准差分别为3.6mm、5.3mm、5.4mm、0.7度和1.4度。T型杆和EFID之间的3D摆位重复性无统计学显著差异(p = 0.22)。对使用两种固定装置治疗的患者进行亚组分析也未发现差异。两种固定装置均未发现从模拟机到治疗单元的一致系统误差。
虽然胸部放疗的最佳固定技术和患者体位尚未确定,但本研究表明T型杆在摆位重复性方面与EFID相当。鉴于T型杆的固有优势,它已成为我们机构的标准固定装置。观察到的两种固定装置在射野定位中的位移范围表明,1cm(摆位误差的两个标准差[SD])将是肺癌放射治疗中允许摆位变化的更合适边界。