Reutern G M, Büdingen H J, Hennerici M, Freund H J
Arch Psychiatr Nervenkr (1970). 1976 Oct 28;222(2-3):191-207. doi: 10.1007/BF02206617.
The Doppler examination of the carotid arteries in the neck was compared with the results of the examination of the supratrochlear artery and of the angiographic controls in 76 patients. The results of this study indicate that the insonation of the carotid arteries considerably improves the diagnostic possibilities offered by the examination of the supratrochlear artery alone. There are three main advantages: 1. The reliability of the Doppler examination is improved. The insonation of the carotid arteries of the 76 angiographically controlled patients including stenoses from 50% onward gave only one false-positive and one false-negative result. The percentage of correct diagnoses thereby attained was 97%. 2. The insonation of the carotid arteries permits reliable differentiation between stenoses and occlusions. This is not possible by the insonation of the supratrochlear artery alone. There was only one case in which the distinction was wrong. 3. The localization of the pathologic process was possible in all cases. These conclusions are only valid for the detection of occlusions and stenoses of more than 50%. Low degree stenoses cannot be detected by this method.
对76例患者颈部颈动脉的多普勒检查结果与滑车上动脉检查结果及血管造影对照结果进行了比较。本研究结果表明,对颈动脉进行超声检查大大提高了仅通过滑车上动脉检查所提供的诊断可能性。主要有三个优点:1. 提高了多普勒检查的可靠性。对76例经血管造影对照的患者的颈动脉进行超声检查,包括50%及以上的狭窄,仅出现1例假阳性和1例假阴性结果。由此获得的正确诊断百分比为97%。2. 对颈动脉进行超声检查可可靠地区分狭窄和闭塞。仅通过滑车上动脉超声检查无法做到这一点。只有1例区分错误。3. 在所有病例中都能够对病变过程进行定位。这些结论仅适用于检测50%以上的闭塞和狭窄。低度狭窄无法通过该方法检测到。