Green S, Palethorpe J E, Peach D, Bradley D A
RRPPS, Edgbaston, Birmingham, U.K.
Appl Radiat Isot. 1999 Jan;50(1):137-52. doi: 10.1016/s0969-8043(98)00030-x.
Experiences of the Regional Radiation Physics and Protection Service (RRPPS) in performance assessment of diagnostic X-ray QA instrumentation and on-patient dosemeters are recounted. Issues relating to the provision of realistic and reproducible reference conditions for calibrated X-irradiations are considered and summary statistics from test measurements of dose and kVp meters are provided. For both dose and kVp meters it is indicated that as many as 25% of instruments used in routine use in the U.K. may require some adjustment before they can truly be said to be performing as the manufacturer intended. Results from intercomparison exercises for patient dosimetry services are also discussed. It is apparent that, for those centres participating in the exercise, dose assessments are generally being obtained to within a bias and a relative standard deviation of less then 10%.