Analysis of a mixture of polychlorinated biphenyls and chlorinated pesticides in human serum by column fractionation and dual-column capillary gas chromatography with electron capture detection.

作者信息

Najam A R, Korver M P, Williams C C, Burse V W, Needham L L

机构信息

Centers for Disease Control and Prevention, National Center for Environmental Health, Atlanta, GA 30341-3724, USA.

出版信息

J AOAC Int. 1999 Jan-Feb;82(1):177-85.

Abstract

An analytical method is presented for precise identification and quantitation of 29 specific polychlorinated biphenyl (PCB) congeners and 15 chlorinated pesticides in human serum. Analyte surrogates PCB 30, PCB 204, 2,2',4,4',5,5'-hexabromo-biphenyl, perthane, alpha-hexachlorocyclohexane, and dichlorobenzophenone were added to each sample. The serum was extracted with an organic solvent and separated by adsorption chromatography into 3 elution fractions for high-resolution gas chromatographic analysis. Each fraction was analyzed by dual-column capillary chromatography followed by electron capture detection. Two capillary columns, DB-5 and DB-1701, with different polarities were used to increase selectivity for each analyte. Quantitation was performed by selecting 2 sets of calibration standard mixtures and 1,2-dichloronaphthalene as an internal standard. Mean recoveries ranged from 39 to 126% for selected analytes and from 31 to 88% for surrogates. Detection limits for specific congeners and pesticides are reported. Typical chromatographic profiles of calibration standard mixtures, as well as a human sample, are illustrated. Verification of each analyte is assessed, and results of analyses of selected human samples and quality control criteria used to ensure data validity also are presented.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验