Comparison of different biomaterials for glaucoma drainage devices.

作者信息

Ayyala R S, Harman L E, Michelini-Norris B, Ondrovic L E, Haller E, Margo C E, Stevens S X

机构信息

Department of Ophthalmology, University of South Florida, Tampa, USA.

出版信息

Arch Ophthalmol. 1999 Feb;117(2):233-6. doi: 10.1001/archopht.117.2.233.

Abstract

OBJECTIVES

To compare the inflammatory reaction associated with the insertion of silicone and polypropylene endplates and endplates made of a new biocompatible polymer, Vivathane, in the rabbit subconjunctival space.

METHODS

Similar-sized endplates made of 3 different biomaterials were sutured to the sclera in the superotemporal quadrant of the rabbit eye. Thirty eyes of 15 albino New Zealand rabbits were randomly assigned to the 3 groups. Conjunctival vascular hyperemia was graded in a masked fashion among groups. At the end of 3 weeks, the enucleated eyes were examined histologically and using scanning electron microscopy.

RESULTS

Polypropylene and Vivathane were associated with significantly more inflammation in clinical observations and based on histological grading. Silicone was associated with the least amount of inflammation. Three polypropylene and 1 Vivathane plate were extruded between the second and third week.

CONCLUSIONS

Silicone is the most inert of the 3 materials tested. Inflammation associated with biomaterials may contribute to the failure of the glaucoma drainage devices.

CLINICAL RELEVANCE

Bleb inflammation may be related to the biomaterial being used as the endplate. Endplates should be handled carefully during surgery to avoid creating rough spots.