非接触式角膜麻醉计的可靠性及其与科谢-博内麻醉计的比较。

Reliability of the non-contact corneal aesthesiometer and its comparison with the Cochet-Bonnet aesthesiometer.

作者信息

Murphy P J, Lawrenson J G, Patel S, Marshall J

机构信息

Department of Vision Sciences, Glasgow Caledonian University, UK.

出版信息

Ophthalmic Physiol Opt. 1998 Nov;18(6):532-9.

Abstract

AIMS

To determine the repeatability of measurement of the corneal sensation threshold using the Non-Contact Corneal Aesthesiometer (NCCA); to evaluate the effect of the air-jet diameter and stimulus duration on the threshold; and to compare the sensation thresholds found with the NCCA and with the Cochet-Bonnet Aesthesiometer (C-BA), on the same group of normal, human subjects.

METHODS

For all experiments, only the central corneal sensation thresholds were measured. (1) NCCA Repeatability; 17 subjects tested at the same time on five separate days. (2) NCCA Air-Jet Diameter; 15 subjects tested using four different air-jets: 0.35, 0.5, 1 and 2 mm. (3) NCCA Stimulus Duration; 15 subjects tested using three stimulus durations: 0.5, 0.9 and 1.5 sec. (4) NCCA and C-BA; 43 subjects tested using each instrument in turn.

RESULTS

(1) No significant differences in mean sensation threshold were detected across the five days of testing, or between individual days. (2) 0.5 and 1 mm air-jets produce least variability in mean sensation threshold. (3) No significant difference in mean sensation thresholds was found between the three test durations. (4) For each subject, there was no correlation between each instrument's mean sensation threshold.

CONCLUSIONS

(1) The NCCA was able to assess the corneal sensation threshold in an accurate and repeatable manner. (2) The optimum test parameters were found using the 0.5 mm air-jet, over a stimulus duration of 0.9 sec. (3) The C-BA has serious deficiencies in its design which limit its ability to accurately measure corneal sensitivity at low stimulus thresholds.

摘要

目的

使用非接触式角膜感觉计(NCCA)确定角膜感觉阈值测量的可重复性;评估喷气直径和刺激持续时间对阈值的影响;并比较在同一组正常人类受试者中使用NCCA和Cochet-Bonnet感觉计(C-BA)测得的感觉阈值。

方法

在所有实验中,仅测量中央角膜感觉阈值。(1)NCCA可重复性;17名受试者在五个不同日期同时进行测试。(2)NCCA喷气直径;15名受试者使用四种不同的喷气:0.35、0.5、1和2毫米进行测试。(3)NCCA刺激持续时间;15名受试者使用三种刺激持续时间:0.5、0.9和1.5秒进行测试。(4)NCCA和C-BA;43名受试者依次使用每种仪器进行测试。

结果

(1)在五天的测试中或不同日期之间,未检测到平均感觉阈值有显著差异。(2)0.5和1毫米的喷气在平均感觉阈值方面产生的变异性最小。(3)在三种测试持续时间之间,未发现平均感觉阈值有显著差异。(4)对于每个受试者,每种仪器的平均感觉阈值之间没有相关性。

结论

(1)NCCA能够以准确且可重复的方式评估角膜感觉阈值。(2)最佳测试参数是使用0.5毫米的喷气,刺激持续时间为0.9秒。(3)C-BA在设计上存在严重缺陷,限制了其在低刺激阈值下准确测量角膜敏感性的能力。