Murphy P J, Patel S, Marshall J
Department of Vision Sciences, Glasgow Caledonian University, UK.
Ophthalmic Physiol Opt. 1996 Mar;16(2):101-7.
A novel method for testing the corneal nerve function, through non-invasive measurement of corneal sensitivity, is presented. The method of using a controlled pulse of air to stimulate the corneal surface is described. Technical information of the aesthesiometer's construction is given. Examples of the ability of the instrument to accurately locate the corneal sensitivity threshold using either a method of constant stimuli, or a method of limits, is described for 14 normal, non-lens wearing eyes. Measurements are made in millibars of air pressure required. Mean normal values found were 0.342 +/- 0.068 millibars. Discussion is made of the instrument's mode of action, its advantages over other invasive instruments, and potential clinical application.
本文介绍了一种通过非侵入性测量角膜敏感度来测试角膜神经功能的新方法。描述了使用可控空气脉冲刺激角膜表面的方法。给出了麻醉计构造的技术信息。针对14只正常、未佩戴镜片的眼睛,描述了该仪器使用恒定刺激法或极限法准确定位角膜敏感度阈值的能力示例。测量以所需气压毫巴为单位进行。发现的正常平均值为0.342±0.068毫巴。讨论了该仪器的作用方式、与其他侵入性仪器相比的优势以及潜在的临床应用。