放疗剂量水平下TLD-100微型立方体的再现性研究。

Reproducibility study of TLD-100 micro-cubes at radiotherapy dose level.

作者信息

da Rosa L A, Regulla D F, Fill U A

机构信息

Instituto de Engenharia Nuclear, IEN, CNEN/RJ, Brazil.

出版信息

Appl Radiat Isot. 1999 Mar;50(3):573-7. doi: 10.1016/s0969-8043(98)00068-2.

Abstract

The precision of the thermoluminescent response of Harshaw micro-cube dosimeters (TLD-100), evaluated in both Harshaw thermoluminescent readers 5500 and 3500, for 1 Gy dose value, was investigated. The mean reproducibility for micro-cubes, pre-readout annealed at 100 degrees C for 15 min, evaluated with the manual planchet reader 3500, is 0.61% (1 standard deviation). When micro-cubes are evaluated with the automated hot-gas reader 5500, reproducibility values are undoubtedly worse, mean reproducibility for numerically stabilised dosimeters being equal to 3.27% (1 standard deviation). These results indicate that the reader model 5500, or, at least, the instrument used for the present measurements, is not adequate for micro-cube evaluation, if precise and accurate dosimetry is required. The difference in precision is apparently due to geometry inconsistencies in the orientation of the imperfect micro-cube faces during readout, requiring careful and manual reproducible arrangement of the selected micro-cube faces in contact with the manual reader planchet.

摘要

研究了在哈肖5500型和3500型热释光读数器中评估的哈肖微立方剂量计(TLD - 100)对1 Gy剂量值的热释光响应精度。对于在100摄氏度下预读出退火15分钟的微立方,使用手动托盘读数器3500评估的平均再现性为0.61%(1个标准差)。当使用自动热气体读数器5500评估微立方时,再现性值无疑更差,数值稳定剂量计的平均再现性等于3.27%(1个标准差)。这些结果表明,如果需要精确准确的剂量测定,5500型读数器,或者至少是用于当前测量的仪器,不适合用于微立方评估。精度差异显然是由于在读数过程中不完美微立方面的取向存在几何不一致,这需要仔细且手动可重复地安排所选微立方面与手动读数器托盘接触。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验