O'Doherty J
Gammaster Ireland Ltd, Westport, Co. Mayo, Ireland.
Med Device Technol. 1997 Jul-Aug;8(6):16-20, 22-3.
There are three aspects that manufacturers must address when validating radiation sterilization. These are the minimum dose that ensures sterility, the maximum dose at which radiation affects the functionality of the device, and the dose distribution throughout the product. This article defines a validation protocol that satisfies current radiation sterilization standards.
制造商在验证辐射灭菌时必须考虑三个方面。这些方面包括确保无菌所需的最小剂量、辐射影响设备功能的最大剂量以及产品中的剂量分布。本文定义了一个符合当前辐射灭菌标准的验证方案。