Wochos J F, Fullerton G D, DeWerd L A
AJR Am J Roentgenol. 1978 Oct;131(4):617-9. doi: 10.2214/ajr.131.4.617.
A method of providing in vivo measurement of entrance skin exposure for mammography using mailed thermoluminescent dosimeters has been developed. The dosimeters are calibrated to 137Cs (662 keV), and a correction is applied to account for energy response. A mailed dosimeter packet which can be used to measure the half-value layer at mammography energies is also described. The results for both systems agree well (to within 10%) with careful ion chamber determinations. These dosimeter measurements provide a simple. Inexpensive means to reliably determine exposure and to monitor machine parameters on a periodic basis.