[医用X线摄影胶片和荧光摄影胶片的感光特性。与胶片组合使用的增感屏的感光度]

[Sensitometric characteristics of medical roentgenographic and fluorographic films. Sensitivity of the amplifying screens combined with a film].

作者信息

Gurvich A M, Erofeeva N D, Katomina R V, Maslov L A

出版信息

Med Tekh. 1976 Mar-Apr(2):17-22.

PMID:1030760
Abstract

Factors affecting the degree of sensitivity of roentgenographic and radiofluorographic films and of their combinations with fluorescent screens are considered. A method of films exposure with a roentgenosensitometer permitting it to determine the sensitivity in conditions closely approaching the ones used in practice and corresponding to the minimum of the isoopaque is suggested. A way for determining the sensitivity of films in m2-mJ1 and lx-1-c-1 is indicated. Numerical values for the sensitivity of the films and their combinations with the screens are given. It is shown that with Eef = 38 keV, T = 20 degrees C, exposure time of 0.4--1 sec and optical density 1.00 over general focusing the lower limit of the X-ray RM-1film sensitivity to the light of CaWO4-screens is 38 m2-mJ-1 and of the fluorographic film Pphi-3 to the light of the Zn0-68Cd0-32S--Ag-screens -- 14.6 m2-mJ-1 or 30 lx-1-c-1.

摘要

考虑了影响X射线摄影胶片、放射性荧光摄影胶片及其与荧光屏组合的敏感度的因素。提出了一种使用X射线感光计进行胶片曝光的方法,该方法能够在接近实际使用条件且对应于等密度最小值的情况下测定敏感度。指出了测定胶片在m²-mJ⁻¹和lx⁻¹-c⁻¹下敏感度的方法。给出了胶片及其与屏幕组合的敏感度数值。结果表明,在Eef = 38 keV、T = 20℃、曝光时间为0.4 - 1秒且光学密度为1.00的情况下,相对于一般聚焦,X射线RM - 1胶片对CaWO₄屏幕光的敏感度下限为38 m²-mJ⁻¹,而荧光摄影胶片Pphi - 3对Zn0 - 68Cd0 - 32S - Ag屏幕光的敏感度下限为14.6 m²-mJ⁻¹或30 lx⁻¹-c⁻¹。