Maslov L A, Gurvich A M, Chikirdin E G, Il'ina M A, Popova T A
Med Tekh. 1988 Sep-Oct(5):36-41.
Three methods of roentgen sensimetry have been compared: the wedge method based on using the stepped wedge, the Moscow Radiation and Roentgenology Research Institute method consisting in the time-modulation of the exposure by means of a lead diaphragm and revolving cassette with screens and film, and the method, according to which the dose is modulated by changing the distance from the source of irradiation to the cassette. The exposure conditions have been selected in order to achieve the best coordination of the results obtained by all the three methods. The minimal sensitivity of the combination of the medium amplification screens of ZY-B2A type with film PM-1 is about 1000P-1. For a more complete characteristic of the important properties of the roentgenographic films it is necessary in addition to determining the sensitivity, gamma and medium gradient to measure the parameters determining the graininess of the film and the transmission of the screen light by its emulsion layer.
已对三种X射线感光度测量方法进行了比较:基于使用阶梯楔形板的楔形法、莫斯科辐射与放射学研究所的方法(该方法通过铅制光阑和带有增感屏及胶片的旋转暗盒对曝光进行时间调制)以及通过改变从辐射源到暗盒的距离来调制剂量的方法。已选择曝光条件,以便使所有这三种方法所获得的结果达到最佳协调。ZY - B2A型中倍率增感屏与PM - 1胶片组合的最小感光度约为1000P - 1。为了更全面地了解X射线胶片的重要特性,除了确定感光度、伽马值和中梯度外,还需要测量决定胶片颗粒度以及增感屏乳剂层对光线透射率的参数。