Damji K F, Shah K C, Rock W J, Bains H S, Hodge W G
University Of Ottawa Eye Institute, Ottawa Hospital, Ottawa, Ontario, Canada.
Br J Ophthalmol. 1999 Jun;83(6):718-22. doi: 10.1136/bjo.83.6.718.
To compare the effectiveness of selective laser trabeculoplasty (SLT, a 532 nm Nd:YAG laser) with argon laser trabeculoplasty (ALT) in lowering the intraocular pressure (IOP) in patients with medically uncontrolled open angle glaucoma.
A prospective randomised clinical trial was designed. Patients were randomised to treatment with either SLT or ALT and were evaluated at 1 hour, 1 week, 1, 3, and 6 months post-laser.
There were 18 eyes in each group. Baseline characteristics were similar in both groups. In the SLT group the mean IOP at baseline, 1, 3, and 6 months was 22.8 (SD 3.0), 20.1 (4.6), 19.3 (6.0), and 17.8 (4.8) mm Hg, respectively. In the ALT group, the mean IOP at baseline, 1, 3, and 6 months was 22.5 (3.6), 19.5 (4.7), 19.6 (5.6), and 17.7 (3.3) mm Hg, respectively. There was a greater anterior chamber reaction, 1 hour after SLT v ALT (p< 0.01). Patients with previous failed ALT had a better reduction in IOP with SLT than with repeat ALT (6.8 (2. 4) v 3.6 (1.8) mm Hg; p = 0.01).
SLT appears to be equivalent to ALT in lowering IOP during the first 6 months after treatment. There is a slightly greater anterior chamber reaction 1 hour after SLT. Patients with previous failed ALT had a significantly greater drop in IOP when treated with SLT v ALT. These results need to be confirmed with a larger sample size.
比较选择性激光小梁成形术(SLT,一种532纳米钕:钇铝石榴石激光)与氩激光小梁成形术(ALT)降低药物治疗无法控制的开角型青光眼患者眼压(IOP)的效果。
设计一项前瞻性随机临床试验。患者被随机分为接受SLT或ALT治疗,并在激光治疗后1小时、1周、1个月、3个月和6个月进行评估。
每组有18只眼。两组的基线特征相似。SLT组在基线、1个月、3个月和6个月时的平均眼压分别为22.8(标准差3.0)、20.1(4.6)、19.3(6.0)和17.8(4.8)毫米汞柱。ALT组在基线、1个月、3个月和6个月时的平均眼压分别为22.5(3.6)、19.5(4.7)、19.6(5.6)和17.7(3.3)毫米汞柱。SLT治疗后1小时的前房反应比ALT更强烈(p<0.01)。既往ALT治疗失败的患者接受SLT治疗后的眼压降低幅度比再次接受ALT治疗更大(6.8(2.4)对3.6(1.8)毫米汞柱;p = 0.01)。
在治疗后的前6个月,SLT在降低眼压方面似乎与ALT相当。SLT治疗后1小时的前房反应略大。既往ALT治疗失败的患者接受SLT治疗时的眼压下降幅度明显大于接受ALT治疗。这些结果需要更大样本量来证实。