Kouri T T, Kähkönen U, Malminiemi K, Vuento R, Rowan R M
Department of Clinical Chemistry, Tampere University Hospital, Finland.
Am J Clin Pathol. 1999 Jul;112(1):25-35. doi: 10.1093/ajcp/112.1.25.
We evaluated the Sysmex UF-100 urine flow cytometer (TOA Medical Electronics, Kobe, Japan) with 269 uncentrifuged urine specimens by comparing it with Sternheimer staining and particle counting in 1-microL disposable chambers with both brightfield and phase-contrast microscopy (the reference method). Results of routine test strip analysis, sediment microscopy (182 specimens), and bacterial culture (204 specimens) were also available. Detection of urinary WBCs and RBCs was highly reliable with the UF-100 compared with manual chamber counting (r = .98 and .88, respectively). Identification of bacteria was equal to that with visual microscopy of uncentrifuged specimens; sensitivity was 55%, and specificity 90%, compared with bacterial cultures at a cutoff of > 10(3) colony-forming units per milliliter. Renal damage was difficult to evaluate even with manual methods because of the low counts of renal tubular cells and casts; with standard manual Sternheimer-stained sediment analysis, sensitivity was 65% to 69% and specificity 66% to 91%, compared with the uncentrifuged chamber method at a cutoff of 3 and 10 particles per microliter, respectively. Renal damage was demonstrated with the UF-100 with a sensitivity of 26% to 69% and specificity 92% to 94%, compared with chamber counts. Automated urinalysis with the UF-100 urine flow cytometer offers considerable savings in time and labor. When high sensitivity is needed, visual microscopic review should be performed to detect renal disease.
我们使用Sysmex UF - 100尿流式细胞仪(日本神户东亚医用电子株式会社)对269份未离心的尿液标本进行了评估,将其与Sternheimer染色以及在1微升一次性计数板中进行明场和相差显微镜下的颗粒计数(参考方法)进行比较。常规试纸条分析、沉渣显微镜检查(182份标本)和细菌培养(204份标本)的结果也可获得。与手工计数板计数相比,UF - 100检测尿白细胞和红细胞的可靠性很高(r分别为0.98和0.88)。细菌鉴定与未离心标本的目视显微镜检查相当;与每毫升>10³菌落形成单位的细菌培养相比,敏感性为55%,特异性为90%。由于肾小管细胞和管型计数较低,即使采用手工方法也难以评估肾损伤;与未离心计数板法相比,采用标准手工Sternheimer染色沉渣分析,在每微升3个和10个颗粒的临界值时,敏感性为65%至69%,特异性为66%至91%。与计数板计数相比,UF - 100显示肾损伤的敏感性为26%至69%,特异性为92%至94%。使用UF - 100尿流式细胞仪进行自动化尿液分析可节省大量时间和人力。当需要高敏感性时,应进行目视显微镜复查以检测肾脏疾病。