Koken Tulay, Aktepe Orhan C, Serteser Mustafa, Samli Murat, Kahraman Ahmet, Dogan Nurhan
Department of Biochemistry, School of Medicine, Afyon Kocatepe, Turkey.
Int Urol Nephrol. 2002;34(2):175-8. doi: 10.1023/a:1023292113462.
Leucocytes and bacteria in urine are the important indices for diagnosing urinary tract infections. Urine specimens from 260 patients were examined for leucocytes and bacteria with the UF-100 flow cytometer (Sysmex Corporation, Japan). The results obtained with the UF-100 were compared with those obtained by bacteriologic cultures of the same urine samples. Bacteriologic samples were found to be positive in 48 urine samples. The best detected cut-off value for bacteria is 3800/microL (senstivity = 73%, specificity = 78%) and 25/microL for leucocytes (sensitivity = 73%, specificity = 92%). In addition, comparison of leucocyte count by manually and with UF-100 shows good correlation (r = 0.91). A flow cytometric urinalysis analyzer operates in a time saving manner, increases throughput and decrease microscopy review rates. It also provides a reliable information to clinicians prior to bacteriologic culture results, in terms of leucocyte and bacteria count.
尿液中的白细胞和细菌是诊断尿路感染的重要指标。使用UF-100流式细胞仪(日本Sysmex公司)对260例患者的尿液标本进行白细胞和细菌检测。将UF-100检测结果与相同尿液样本的细菌培养结果进行比较。细菌培养样本在48份尿液样本中呈阳性。细菌的最佳检测临界值为3800/微升(敏感性=73%,特异性=78%),白细胞为25/微升(敏感性=73%,特异性=92%)。此外,手工计数白细胞与UF-100计数白细胞的比较显示出良好的相关性(r=0.91)。流式细胞尿分析仪操作省时,提高了通量并降低了显微镜复查率。在细菌培养结果出来之前,它还能就白细胞和细菌计数为临床医生提供可靠信息。