确定尿流式细胞仪(UF-100)检测尿路感染时白细胞和细菌的临界值。

Determination of cut-off values for leucocytes and bacteria for urine flow cytometer (UF-100) in urinary tract infections.

作者信息

Koken Tulay, Aktepe Orhan C, Serteser Mustafa, Samli Murat, Kahraman Ahmet, Dogan Nurhan

机构信息

Department of Biochemistry, School of Medicine, Afyon Kocatepe, Turkey.

出版信息

Int Urol Nephrol. 2002;34(2):175-8. doi: 10.1023/a:1023292113462.

Abstract

Leucocytes and bacteria in urine are the important indices for diagnosing urinary tract infections. Urine specimens from 260 patients were examined for leucocytes and bacteria with the UF-100 flow cytometer (Sysmex Corporation, Japan). The results obtained with the UF-100 were compared with those obtained by bacteriologic cultures of the same urine samples. Bacteriologic samples were found to be positive in 48 urine samples. The best detected cut-off value for bacteria is 3800/microL (senstivity = 73%, specificity = 78%) and 25/microL for leucocytes (sensitivity = 73%, specificity = 92%). In addition, comparison of leucocyte count by manually and with UF-100 shows good correlation (r = 0.91). A flow cytometric urinalysis analyzer operates in a time saving manner, increases throughput and decrease microscopy review rates. It also provides a reliable information to clinicians prior to bacteriologic culture results, in terms of leucocyte and bacteria count.

摘要

尿液中的白细胞和细菌是诊断尿路感染的重要指标。使用UF-100流式细胞仪(日本Sysmex公司)对260例患者的尿液标本进行白细胞和细菌检测。将UF-100检测结果与相同尿液样本的细菌培养结果进行比较。细菌培养样本在48份尿液样本中呈阳性。细菌的最佳检测临界值为3800/微升(敏感性=73%,特异性=78%),白细胞为25/微升(敏感性=73%,特异性=92%)。此外,手工计数白细胞与UF-100计数白细胞的比较显示出良好的相关性(r=0.91)。流式细胞尿分析仪操作省时,提高了通量并降低了显微镜复查率。在细菌培养结果出来之前,它还能就白细胞和细菌计数为临床医生提供可靠信息。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验