The sensitivity of amino acid analysis by electrospray-tandem mass spectrometry (ESI-TMS) in differential diagnosis of the different forms of hyperphenylalaniemias (HPA) was studied in a total number of 78 dried blood samples from neonates with increased phenylalanine (Phe) concentrations (> 150 micromol/l). Simultaneous quantitative analysis of Phe and tyrosine (Tyr) were performed by ESI-TMS. By means of the Phe/Tyr-ratio it was possible to differentiate between phenylketonuria (PKU; n = 23) and nonPKU-HPA patients (n = 35; P < 0.001) with a sensitivity of 100%. Neonates with transient HPA (n = 9) showed a lower Phe/Tyr-ratio compared to the nonPKU-HPA patients (P < 0.001). Amino acid analysis by ESI-TMS represents a powerful diagnostic tool in differential diagnosis of HPA and is an appropriate technique in the setting of neonatal screening based on dried blood spots.
在78份来自苯丙氨酸(Phe)浓度升高(>150微摩尔/升)的新生儿的干血样本中,研究了电喷雾串联质谱法(ESI-TMS)进行氨基酸分析在不同形式高苯丙氨酸血症(HPA)鉴别诊断中的敏感性。通过ESI-TMS对Phe和酪氨酸(Tyr)进行同步定量分析。借助Phe/Tyr比值,能够以100%的敏感性区分苯丙酮尿症(PKU;n = 23)和非PKU-HPA患者(n = 35;P < 0.001)。与非PKU-HPA患者相比,短暂性HPA新生儿(n = 9)的Phe/Tyr比值更低(P < 0.001)。ESI-TMS进行的氨基酸分析是HPA鉴别诊断中的一种强大诊断工具,也是基于干血斑进行新生儿筛查的合适技术。