Krupinski E, Roehrig H, Furukawa T
Department of Radiology, University of Arizona, Tucson 85724, USA.
Acad Radiol. 1999 Jul;6(7):411-8. doi: 10.1016/s1076-6332(99)80191-3.
The purpose of this study was to measure the influence of display luminance on detection performance and visual search behavior. The results of the study should be helpful in establishing minimally acceptable display conditions for viewing radiographs on cathode-ray tube (CRT) monitors.
Two groups of six radiologists each viewed 50 pairs of mammograms. One group viewed film images on a standard mammographic view box; the other viewed images on a high-resolution CRT monitor. Two luminance levels were studied for each display type. Observers reported on the presence or absence of masses or microcalcification clusters and on their confidence in that decision. Confidence data were analyzed by using alternative free-response receiver operating characteristic (AFROC) techniques. Eye position also was recorded as observers viewed the images.
For both the film and monitor studies, detection performance (AFROC area under the curve) was not affected significantly by display luminance, but search behavior was. Total viewing and decision dwell times were shorter with the higher-luminance displays, especially for true-negative decisions. Significantly more fixation clusters were generated during the search of lesion-free than of lesion-containing images with the lower-luminance displays.
Display luminance affects visual search performance with both film and monitor displays without affecting detection performance significantly. Higher-luminance displays yield more efficient search performance. The true-negative dwell times and number of clusters are suggestive that lower-luminance levels prolong the search and recognition of normal, lesion-free areas compared with lesion-containing areas.
本研究旨在测量显示亮度对检测性能和视觉搜索行为的影响。该研究结果应有助于确定在阴极射线管(CRT)显示器上查看X光片时的最低可接受显示条件。
两组放射科医生,每组六人,分别查看50对乳房X光片。一组在标准乳房X光片观片箱上查看胶片图像;另一组在高分辨率CRT显示器上查看图像。对每种显示类型研究了两种亮度水平。观察者报告是否存在肿块或微钙化簇以及他们对该判断的信心。使用替代自由反应接受者操作特征(AFROC)技术分析信心数据。在观察者查看图像时也记录眼睛位置。
对于胶片和显示器研究,检测性能(曲线下AFROC面积)不受显示亮度的显著影响,但搜索行为受影响。高亮度显示器的总查看和决策停留时间较短,尤其是对于真阴性决策。在查看无病变图像时,低亮度显示器产生的注视簇明显多于查看有病变图像时。
显示亮度对胶片和显示器显示的视觉搜索性能有影响,但对检测性能无显著影响。高亮度显示器产生更高效的搜索性能。真阴性停留时间和簇的数量表明,与有病变区域相比,低亮度水平会延长对正常、无病变区域的搜索和识别时间。