Schulz R J
Med Phys. 1978 Nov-Dec;5(6):525-31. doi: 10.1118/1.594443.
The photoneutrons produced by high-energy x-ray machines can diffuse through the mazes usually employed at the treatment-room entrance and readily penetrate the lead-lined doors used for x-ray shielding. The measurement of these neutrons in the presence of x-rays and the determination of dose equivalent poses a problem for which there is currently no standard method of solution. In order to separate x-ray dose from neutron dose, the author employed an ionization chamber alternately filled with argon or propane. The response characteristics of this chamber to x-ray and neutrons are described. Quality factors were determined from a calculated neutron spectrum. As a result of these measurements, a 10-in. polyethylene door was added to the entranceway of a 25-MV linear accelerator.