上颌种植体切割扭矩与共振频率测量的比较。一项为期20个月的临床研究。

A comparison between cutting torque and resonance frequency measurements of maxillary implants. A 20-month clinical study.

作者信息

Friberg B, Sennerby L, Meredith N, Lekholm U

机构信息

The Brånemark Clinic, Faculty of Odontology, Department of Biomaterials/Handicap Research, Göteborg University, Sweden.

出版信息

Int J Oral Maxillofac Surg. 1999 Aug;28(4):297-303.

Abstract

Oral implant treatment ad modum Brånemark was undertaken in nine patients with edentulous maxillae. Cutting torque measurements and resonance frequency analyses (RFA) were conducted at implant placement and the corresponding values were subjected to correlation analyses. The implants were also evaluated with RFA at abutment connection and at one-year follow-up in order to identify possible changes in implant stability. A total of 61 implants were inserted, of which 49 were of the Mk II self-tapping type. Two implants were lost during the study period. The cumulative torque was presented as a mean value for the upper/crestal, the middle and the lower/apical third of the implant site respectively, as well as an overall value for the whole site. The highest correlation (r = 0.84, P<0.05) was found when comparing the mean torque values of the upper/crestal portion with the resonance frequency values at implant placement. The Mk II implant sites were divided into three groups based on the values of the cutting torque, i.e. soft (group 1), medium (group 2) and dense bone (group 3). The mean value of each group was plotted against the corresponding mean value of resonance frequency measured at implant insertion. Statistical analysis showed significant differences in resonance frequency at implant insertion between groups 1 and 2 (P = 0.047) and between groups 1 and 3 (P = 0.002). When repeating the resonance frequency analyses at second stage surgery and at one-year follow-up, no significant differences were detected between any of the groups. It was shown that the stability of implants placed in softer bone seemed to "catch up" over time with more dense bone sites.

摘要

对9例上颌无牙患者采用了类似Brånemark系统的口腔种植治疗。在种植体植入时进行了切割扭矩测量和共振频率分析(RFA),并对相应数值进行了相关性分析。为了确定种植体稳定性的可能变化,还在基台连接时和随访1年时用RFA对种植体进行了评估。共植入61枚种植体,其中49枚为Mk II自攻型。在研究期间有2枚种植体丢失。累积扭矩分别以上颌/嵴顶、中部和下颌/根尖三分之一处种植部位的平均值以及整个部位的总值表示。比较上颌/嵴顶部分的平均扭矩值与种植体植入时的共振频率值时,发现相关性最高(r = 0.84,P<0.05)。根据切割扭矩值将Mk II种植部位分为三组,即软质骨(第1组)、中等骨密度(第2组)和致密骨(第3组)。将每组的平均值与种植体植入时测得的相应共振频率平均值作图。统计分析显示,第1组和第2组(P = 0.047)以及第1组和第3组(P = 0.002)在种植体植入时的共振频率存在显著差异。在二期手术和随访1年时重复进行共振频率分析时,未发现任何组之间存在显著差异。结果表明,随着时间的推移,植入较软骨质中的种植体的稳定性似乎会赶上骨质更致密的部位。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验