Arcari M, Phillips S D, Gibbs P, Rela S M, Heaton N D
Liver Transplant Surgical Service, Kings College Hospital, London, UK.
Transplantation. 1999 Jul 15;68(1):150-2. doi: 10.1097/00007890-199907150-00027.
After observing micro-bubble activity in the venovenous bypass system during liver transplantation, an experiment was designed to investigate the origin of these bubbles and to define the conditions under which they occurred. Using a Biomedicus constrained vortex pump and a customized circuit design, microbubble activity was measured in saline and blood media during varying pre- and post-head pressures. The data show that air emboli can be generated from this pump and the rate at which they develop is directly related to the pre- and post-head pressure and hematocrit.
在观察肝移植过程中静脉-静脉转流系统中的微泡活动后,设计了一项实验来研究这些气泡的来源,并确定它们产生的条件。使用一台Biomedicus受限涡流泵和定制的回路设计,在不同的前后头压力下,测量盐水和血液介质中的微泡活动。数据表明,这种泵可产生空气栓子,其形成速率与前后头压力和血细胞比容直接相关。