Gryczynski Z, Abugo O O, Lakowicz J R
Center for Fluorescence Spectroscopy, University of Maryland, Baltimore School of Medicine, 725 West Lombard Street, Baltimore, Maryland 21201, USA.
Anal Biochem. 1999 Sep 10;273(2):204-11. doi: 10.1006/abio.1999.4213.
We used a new method, polarization sensing, to monitor the concentration of the fluorophore rhodamine 800 in an intralipid suspension and in chicken tissue. Rhodamine 800 (Rh800) could be excited at 648 nm using a laser pointer. We developed a simple device for measuring the combined emission from a highly polarized reference film and the unpolarized or orthogonally polarized emission of Rh800 from the scattering intralipid or tissue. The concentration of Rh800 in this medium was revealed by large changes in the polarization (P) with values of P ranging from 0.8 to -0.9. It is possible to vary the sensitive Rh800 concentration range by variation of the detected emission wavelengths, orientation of the excitation polarizer, or fluorophore concentration in the reference film. Polarization sensing of fluorophores in tissue requires only steady-state detection, and can be accomplished with simple and/or portable electronics. Such devices may find use in electronic detection of ingested medicines based on transdermal detection of nontoxic long-wavelength fluorophores.
我们使用了一种新方法——偏振传感,来监测脂质乳剂和鸡组织中荧光团罗丹明800的浓度。使用激光指示器可在648nm激发罗丹明800(Rh800)。我们开发了一种简单装置,用于测量来自高偏振参考膜的组合发射以及来自散射脂质乳剂或组织的Rh800的非偏振或正交偏振发射。通过偏振(P)的大幅变化揭示了该介质中Rh800的浓度,P值范围为0.8至 -0.9。通过改变检测发射波长、激发偏振器的方向或参考膜中的荧光团浓度,可以改变敏感的Rh800浓度范围。组织中荧光团的偏振传感仅需要稳态检测,并且可以通过简单和/或便携式电子设备来完成。此类装置可能会用于基于对无毒长波长荧光团的透皮检测来进行摄入药物的电子检测。