Electrocardiographically gated thin-section CT of the lung.

作者信息

Schoepf U J, Becker C R, Bruening R D, Helmberger T, Staebler A, Leimeister P, Reiser M F

机构信息

Department of Diagnostic Radiology, Klinikum Grosshadern, University of Munich, Germany.

出版信息

Radiology. 1999 Sep;212(3):649-54. doi: 10.1148/radiology.212.3.r99se08649.

Abstract

PURPOSE

To determine whether electrocardiographic (ECG) gating improves image quality of thin-section computed tomographic (CT) scans of the lung obtained with a subsecond CT scanner.

MATERIALS AND METHODS

Thin-section CT was performed in 35 patients by using standard techniques. Three additional sections were obtained in each patient with prospective ECG gating at corresponding levels of the paracardiac lung parenchyma. Non-ECG-gated and ECG-gated sections were then rated in blinded fashion by three experienced radiologists for overall image quality, spatial resolution, and diagnostic value and for different types of respiratory and cardiac motion artifacts.

RESULTS

ECG gating helped significantly reduce artifacts caused by cardiac motion (i.e., distortion of pulmonary vessels, double images, or blurring of the cardiac border) (P < .05). ECG gating did not reduce respiratory motion artifacts. In patients with heart rates of less than 76 beats per minute, ECG gating significantly improved overall image quality (P = .041). ECG gating was not perceived to increase the diagnostic value of thin-section CT scans.

CONCLUSION

ECG gating improves image quality of thin-section CT scans of the lung by reducing cardiac motion artifacts that may mimic disease. It must be established whether ECG gating can help increase the diagnostic accuracy of thin-section CT for the evaluation of subtle parenchymal disease.