Madsen E L, Dong F, Frank G R, Garra B S, Wear K A, Wilson T, Zagzebski J A, Miller H L, Shung K K, Wang S H, Feleppa E J, Liu T, O'Brien W D, Topp K A, Sanghvi N T, Zaitsev A V, Hall T J, Fowlkes J B, Kripfgans O D, Miller J G
Medical Physics Department, University of Wisconsin, Madison, USA.
J Ultrasound Med. 1999 Sep;18(9):615-31. doi: 10.7863/jum.1999.18.9.615.
In a study involving 10 different sites, independent results of measurements of ultrasonic properties on equivalent tissue-mimicking samples are reported and compared. The properties measured were propagation speed, attenuation coefficients, and backscatter coefficients. Reasonably good agreement exists for attenuation coefficients, but less satisfactory results were found for propagation speeds. As anticipated, agreement was not impressive in the case of backscatter coefficients. Results for four sites agreed rather well in both absolute values and frequency dependence, and results from other sites were lower by as much as an order of magnitude. The study is valuable for laboratories doing quantitative studies.
在一项涉及10个不同地点的研究中,报告并比较了在等效组织模拟样本上超声特性测量的独立结果。所测量的特性包括传播速度、衰减系数和背向散射系数。衰减系数方面存在相当不错的一致性,但传播速度的结果不太令人满意。正如预期的那样,背向散射系数的一致性并不显著。四个地点的结果在绝对值和频率依赖性方面都相当吻合,而其他地点的结果则低至一个数量级。该研究对进行定量研究的实验室很有价值。