Visser M, Fuerst T, Lang T, Salamone L, Harris T B
Epidemiology, Demography and Biometry Program, National Institute on Aging, Bethesda, Maryland 20892, USA.
J Appl Physiol (1985). 1999 Oct;87(4):1513-20. doi: 10.1152/jappl.1999.87.4.1513.
The aim of the study was to examine the accuracy of fan-beam dual-energy X-ray absorptiometry (DEXA) for measuring total body fat-free mass (FFM) and leg muscle mass (MM) in elderly persons. Participants were 60 men and women aged 70-79 yr and with a body mass index of 17.5-39.8 kg/m(2). FFM and MM at four leg regions were measured by using DEXA (Hologic 4500A, v8.21). A four-compartment body composition model (4C) and multislice computed tomography (CT) of the legs were used as the criterion methods for FFM and MM, respectively. FFM by DEXA was positively associated with FFM by 4C (R(2) = 0.98, SE of estimate = 1.6 kg). FFM by DEXA was higher [53.5 +/- 12.0 (SD) kg] than FFM by 4C (51.6 +/- 11.9 kg; P < 0.001). No association was observed between the difference and the mean of the two methods. MM by DEXA was positively associated with CT at all four leg regions (R(2) = 0.86-0.96). MM by DEXA was higher than by CT in three regions. The results of this study suggest that fan-beam DEXA offers considerable promise for the measurement of total body FFM and leg MM in elderly persons.
本研究的目的是检验扇形束双能X线吸收法(DEXA)测量老年人全身去脂体重(FFM)和腿部肌肉量(MM)的准确性。研究对象为60名年龄在70 - 79岁、体重指数为17.5 - 39.8 kg/m²的男性和女性。使用DEXA(Hologic 4500A,v8.21)测量四个腿部区域的FFM和MM。分别采用四成分身体成分模型(4C)和腿部多层计算机断层扫描(CT)作为FFM和MM的标准测量方法。DEXA测量的FFM与4C测量的FFM呈正相关(R² = 0.98,估计标准误 = 1.6 kg)。DEXA测量的FFM[53.5 ± 12.0(标准差)kg]高于4C测量的FFM(51.6 ± 11.9 kg;P < 0.001)。未观察到两种方法差异与均值之间的关联。DEXA测量的MM与所有四个腿部区域的CT呈正相关(R² = 0.86 - 0.96)。DEXA测量的MM在三个区域高于CT测量的结果。本研究结果表明,扇形束DEXA在测量老年人全身FFM和腿部MM方面具有很大的前景。