与标准方法相比,QDR 4500A双能X线骨密度仪高估了去脂体重。

QDR 4500A DXA overestimates fat-free mass compared with criterion methods.

作者信息

Tylavsky Frances, Lohman Timothy, Blunt Barbara A, Schoeller Dale A, Fuerst Thomas, Cauley Jane A, Nevitt Michael C, Visser Marjolein, Harris Tamara B

机构信息

Department of Preventive Medicine, University of Tennessee Health Science Center, Memphis, Tennessee 38105, USA.

出版信息

J Appl Physiol (1985). 2003 Mar;94(3):959-65. doi: 10.1152/japplphysiol.00732.2002. Epub 2002 Nov 1.

Abstract

This study evaluated the accuracy with which the dual-energy X-ray absorptiometer (Hologic QDR 4500A) measured fat-free mass (FFM), fat mass (FM), and hydration of FFM. In a study of 58 men and women (ages 70-79 yr), the QDR 4500A was found to provide a systematically higher estimate of FFM and lower estimate of FM than a four-component model of body composition. A correction factor from this study was developed and applied to two other samples (n = 13 and 37). We found mean corrected levels of FFM and FM to be equivalent to that obtained by the four-component model or total body water. In addition, the hydration of the corrected FFM was closer to the established hydration level in adult samples and that obtained from the four-component model. These findings suggest that the current calibration of the fan-beam system of the Hologic QDR 4500A provides an overestimate of FFM and underestimate of FM compared with reference methods.

摘要

本研究评估了双能X线吸收仪(Hologic QDR 4500A)测量去脂体重(FFM)、脂肪量(FM)以及FFM水合作用的准确性。在一项针对58名年龄在70至79岁之间的男性和女性的研究中,发现与身体成分的四成分模型相比,QDR 4500A对FFM的估计系统性地偏高,而对FM的估计偏低。根据该研究制定了一个校正因子,并应用于另外两个样本(n = 13和37)。我们发现校正后的FFM和FM平均水平与通过四成分模型或总体水获得的水平相当。此外,校正后FFM的水合作用更接近成人样本中既定的水合水平以及从四成分模型中获得的水平。这些发现表明,与参考方法相比,Hologic QDR 4500A扇形束系统的当前校准高估了FFM且低估了FM。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验